[
Spectrofluorometry ]
Spectrofluorometry
is commonly defined as a type of spectrophotometry deals
with fluorescence, the phenomenon in which light incident
on a material at a given wavelength, usually in the ultraviolet
or visible spectral range, causes that material to radiate
light at longer, less energetic wavelengths. To observe
this effect, spectrofluorometers have dispersive optics,
usually grating monochromators, both between the source
and the sample for the incident light and between the sample
and the detector for the exitent light. This permits the
identification of the wavelengths that excite the sample
and the wavelengths at which that excitation generates the
fluorescence. Obviously, fluorescence can produce spurious
signals in an optical system, so that fluorescent optics
and their excitation and emission spectra should be recognized
and accounted for. On the other hand, the fluorescence phenomenon
exploited by technologies ranging from household cleaning
products to advanced biotechnology.
ODA
owns one spectrofluorometer, the Varian Eclipse, and operates
another, the Spex Fluoro-Max 2, under an agreement with
the University of Arizona. The attributes of these two instruments
are compared in the following tables:
Wavelength
Range
Instrument |
Nanometer
(nm) Range |
Micron
(µm) Range |
Wavenumber
(1/cm) Range |
Varian
Eclipse |
190 |
1,100 |
0.19 |
1.10 |
52,632 |
9,091 |
|
200 |
900 |
0.20 |
0.90 |
50,000 |
11,111 |
Optical
Arrangement
Instrument |
Layout |
Sources |
Sample
Compartment |
Detectors |
Varian
Eclipse |
Source-monochro-
mator-sample-
monochromator-
detector |
Pulsed
xenon arc lamp |
Single-beam,
focused |
Integrating
photomultiplier |
Spex
Fluoro-Max 2 |
Source-monochro-
mator-sample-
monochromator-
detector |
Continuous
xenon arc lamp |
Single-beam,
focused |
Photon
counting
photomultiplier |
Orientation
Both
instruments can accommodate cuvettes with a standard 1 cm
square cross-section. These are illuminated normally on
one face; the emitted light from an adjacent face, that
is, at a 90° angle from the incident beam, is detected.
Both
instruments can also accommodate solid samples. For the
solid sample attachment for the Varian Eclipse the excitation
beam is incident at 15°; the exitent beam is therefore
at 75°. For the Spex Fluoro-Max 2, the excitation beam
is incident at 20°; the exitent beam is therefore at
70°.
Sample
Properties
Solid
Samples
Flat
Specular Optical Elements and Other Materials
These
materials are measured with direct transmittance sample
holders or with the specular reflectance accessories. Ideally,
samples have diameters or square sides of 2.5 to 7.5 cm
[1 to 3 inches], but samples with dimensions as small as
1 mm [0.040 inch] and as large as 0.75 m [30 inches] can
be accommodated with special jigs and holders.
Curved
Specular or Rough Optical Elements, Powders, and other Materials
These
materials are usually measured at the transmittance or reflectance
port of the appropriate integrating sphere. Ideally, samples
have diameters or square sides of 2.5 to 7.5 cm [1 to 3
inches], but samples with dimensions as small as 5 mm [0.040
inch] and as large as 0.75 m [30 inches] can be accommodated
with special jigs and holders. If both reflectance and transmittance
must be measured at the same time ["transflec-tance"],
the sample can be measured at the center of the sphere ["Edwards"
mode], but the dimensions must not exceed 2.5 cm.
For
the Cary 500E [UV-Vis-NIR], two 150-mm diameter integrating
spheres, both made by Labsphere, are now available.
•
The first is known as a "side-looking" sphere;
the entrance and exit ports for the sample and reeference
beams are arrayed along the horizontally oriented "equator"
of the sphere. The access port is at the "top"
pole, and the detector port is at the "bottom"
pole. This means that samples are mounted "vertically"
on the sides of the sphere.
For
T measurements, the sample is mounted in front of the
entrance port. The specular portion can be included
[total T] if the exit port on the opposite side is covered
with a Spectralon™
standard identical to that of the sphere walls, or it
can be excluded [diffuse T] if the exit port is covered
with a black standard or light trap. Samples up to 10
cm across can be accommodated.
For
R measurements, the specular portion of the beam can
be included [total reflectance] or excluded [diffuse
reflectance], depending on whether a small portion of
the sphere wall - which is on a removable plug - in
the specular direction remains in place or is removed.
Samples up to about two meters in maximum dimension
can be accommodated.
For
powder R measurements with this sphere, ODA emplys a
powder cell made by Avian Technologies 2with a quartz
window and a Spectralon™ cavity; other arrangements
can eb made if necessary.
•
The second is known as a "down-looking" sphere;
the entrance and exit ports for the sample beam are arrayed
along the vertically oriented "equator" of the
sphere. The sample beam can be switched between an entrance
port at the side of the sphere or another entrance port
at the "top" pole of the sphere.
For
T measurements, the sample is placed vertically in front
of the entrance port on the side of the sphere. Only
total T measurements can be made, but samples up to
20 cm across can be accommodated, about double the capacity
of the side-looking sphere.
For
R measurements, the sample is placed horizontally against
the port at the "bottom" pole of the sphere,
such that powders and liquids can be accommodated without
the use of a cell. The beam is directed downward by
a mirror outside the entrance port at the top of the
sphere. As with the "side-looking sphere"
the specualr portion of the beam can be included [total
reflectance] or excluded [diffuse reflectance], depending
on whether a small portion of the sphere wall - which
is on a removable plug - in the specular direction remains
in place or is removed.
A
diffuse reflectance accessory involving specular optics
and a upward-looking powder holder is also available.
For
the Nicolet 560 [IR], a 102 -mm diameter sphere is available.
The entrance port for this sphere is on the side of the
sphere, so transmittance samples are mounted vertically.
Because this sphere employs a specular mirror at its center
to deflect the beam along a vertically oriented great circle,
reflectance samples can be mounted at the bottom port, permitting
the measurement of liquid or powder samples.
There
is no integrating sphere capability for the Perkin-Elmer
983G.
Liquid
Samples
For
the UV-Vis-NIR, ODA maintains a variety of UV quartz, glass,
and plastic cuvettes with path lengths ranging from 1 to
50 mm, including matched pairs.
Gas
Samples
ODA
maintains a single gas cell with a 10 cm path length.
Optical
Density
In
general, for transmittances down to about 0.01= 1%, measurements
are made with no reference beam attenuation in a linear,
transmittance mode . For transmittances below 0.01 = 1%,
attenuators are placed in the rear beam and the measurements
are expressed as absorptance, the negative base 10 logarithm
of the transmittance. The limits for the Cary 500E UV-Vis-NIR
SPM are about A = 6.3 in the UV-Vis ranges, and about 4.5
in the NIR range. The limit of the PE 983G IR SPM is about
3 over the MIR & FIR range.
Temperature
& Humidity
Most
measurements are made at ambient temperature [T]. However,
ODA operates a cryostat that can measure transmittance [T]
at T's down to about 85K, approaching liquid nitrogen's
T of 77 K, For these measurements, a diameter of 2.5 cm
is standard; other sizes will incur tooling charges. ODA
has measured T of samples at up to 800 °C in heater
blocks for special projects. Also, T measurements have been
made for fixed levels of humidity in special cells for periods
of up to six weeks. ODA also performs measurement before
and after exposure to environmental tests to identify failure
modes. ODA can subcontract these exposures if requested.
|